Send Message
Good price  online

products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Dental Handpiece Repair
Created with Pixso. Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance

Brand Name: All kinds of Handpiece(Such as NSK, KaVo, W&H, COXO,etc)
MOQ: 1 piece
Price: USD 20~250
Delivery Time: 2-8 working days after receiving payments
Payment Terms: D/A, D/P, T/T, Western Union
Detail Information
Place of Origin:
China
Certification:
ISO
Packaging Details:
box, 1 SET/BOX, Carton
Highlight:

Precision Dental Handpiece Repair

,

Dental Handpiece Repair Systematic

,

Precision Low Speed Handpiece Repair

Product Description

Dental Handpiece Repair Expert: Quality Service at Competitive Prices

 

 

Alongside producing top-tier dental mobile phones, FHJS provides expert third-party services focused on the systematic repair and maintenance of major dental mobile brands. We serve individuals, corporations, and medical facilities. Our dedicated team of technicians possesses deep expertise in both local and global brands, having accumulated significant experience in repairs, maintenance, upgrades, and operational management throughout their careers.

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 0

 

The repair of dental mobile phones involves many aspects. The following are some common faults:

 

‌The needle cannot be clamped or the needle cannot be removed‌
1. The diameter of the tail of the needle is small and does not meet the requirements.
2. Dirt blocks the chuck groove and the chuck cannot be closed.
3. The three-petal spring is worn and broken: replace the three-petal spring.
4. The needle and the clamping mechanism are rusted.

 

 

The mobile phone does not produce water or the water output is small‌:


1. The water path of the mobile phone leather line is blocked.
2. There is dirt in the water spray port of the mobile phone.
3. The internal water path of the mobile phone is blocked.

 

Water leakage at the connection between the handpiece and the leather line or the quick-change connector‌

 

1. The connection between the handpiece and the leather line is not tightened.
2. The four-hole or two-hole sealing gasket at the tail of the handpiece is aging, deformed, or damaged.
3. The front end of the quick-change connector is deformed after being dropped.
4. The small O-ring at the front end of the quick-change connector is severely worn.

 

 The sound is abnormal, the noise becomes louder, the speed is slow, and the cutting force is too small when the handpiece is used‌

 

1. The head of the handpiece is injured and deformed, and the wind wheel has great resistance to operation.
2. The bearing lacks cleaning and lubrication.
3. The bearing is damaged.

 

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 1


In addition, for low-speed dental handpieces, if the engine does not rotate, the straight machine does not rotate when it is connected to the engine, the bending machine slips during operation, or the needle swings violently or does not rotate, etc., corresponding maintenance and processing are also required according to the specific situation.

 

If necessary, please leave a message for inquiry.

 

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 2

 

Good price  online

Products Details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Dental Handpiece Repair
Created with Pixso. Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance

Brand Name: All kinds of Handpiece(Such as NSK, KaVo, W&H, COXO,etc)
MOQ: 1 piece
Price: USD 20~250
Packaging Details: box, 1 SET/BOX, Carton
Payment Terms: D/A, D/P, T/T, Western Union
Detail Information
Place of Origin:
China
Brand Name:
All kinds of Handpiece(Such as NSK, KaVo, W&H, COXO,etc)
Certification:
ISO
Minimum Order Quantity:
1 piece
Price:
USD 20~250
Packaging Details:
box, 1 SET/BOX, Carton
Delivery Time:
2-8 working days after receiving payments
Payment Terms:
D/A, D/P, T/T, Western Union
Highlight:

Precision Dental Handpiece Repair

,

Dental Handpiece Repair Systematic

,

Precision Low Speed Handpiece Repair

Product Description

Dental Handpiece Repair Expert: Quality Service at Competitive Prices

 

 

Alongside producing top-tier dental mobile phones, FHJS provides expert third-party services focused on the systematic repair and maintenance of major dental mobile brands. We serve individuals, corporations, and medical facilities. Our dedicated team of technicians possesses deep expertise in both local and global brands, having accumulated significant experience in repairs, maintenance, upgrades, and operational management throughout their careers.

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 0

 

The repair of dental mobile phones involves many aspects. The following are some common faults:

 

‌The needle cannot be clamped or the needle cannot be removed‌
1. The diameter of the tail of the needle is small and does not meet the requirements.
2. Dirt blocks the chuck groove and the chuck cannot be closed.
3. The three-petal spring is worn and broken: replace the three-petal spring.
4. The needle and the clamping mechanism are rusted.

 

 

The mobile phone does not produce water or the water output is small‌:


1. The water path of the mobile phone leather line is blocked.
2. There is dirt in the water spray port of the mobile phone.
3. The internal water path of the mobile phone is blocked.

 

Water leakage at the connection between the handpiece and the leather line or the quick-change connector‌

 

1. The connection between the handpiece and the leather line is not tightened.
2. The four-hole or two-hole sealing gasket at the tail of the handpiece is aging, deformed, or damaged.
3. The front end of the quick-change connector is deformed after being dropped.
4. The small O-ring at the front end of the quick-change connector is severely worn.

 

 The sound is abnormal, the noise becomes louder, the speed is slow, and the cutting force is too small when the handpiece is used‌

 

1. The head of the handpiece is injured and deformed, and the wind wheel has great resistance to operation.
2. The bearing lacks cleaning and lubrication.
3. The bearing is damaged.

 

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 1


In addition, for low-speed dental handpieces, if the engine does not rotate, the straight machine does not rotate when it is connected to the engine, the bending machine slips during operation, or the needle swings violently or does not rotate, etc., corresponding maintenance and processing are also required according to the specific situation.

 

If necessary, please leave a message for inquiry.

 

 

 

Precision High / Low Speed Dental Handpiece Repair Systematic Repair And Maintenance 2